Thermocouple & RTD batch converter
Turn a whole column of thermocouple voltages (mV) or RTD resistances (Ω) from your DAQ into °C at once. Paste a column copied from Excel, or pick a CSV file.
🔒 Calculated in this browser only — your data never leaves your PC
How it is calculated
- Thermocouples: NIST ITS-90 reference and inverse functions (piecewise polynomials). The cold-junction EMF is added to each reading before solving for temperature. Values outside the table range are flagged.
- RTD: Callendar–Van Dusen equation. Pt100 · Pt500 · Pt1000 use the IEC 60751 coefficients (A 3.9083×10⁻³ · B −5.775×10⁻⁷ · C −4.183×10⁻¹²), including the C term below 0 °C.
- The same coefficients and math the Runplotter program uses to decode thermocouple and RTD channels in NI TDMS files.